Characteristics
- Measured entity
- signal
- Application domain
- monitoring, process
- Measured value
- spectrum
- Configuration
- online
- Technology
- optical
- Other characteristics
- digital, multi-channel, pulse, MCA
Description
Eight independent 32k digital MCA in VME form factor
Perfectly suited for high resolution gamma ray spectroscopy with
single and segmented HPGe, Clover and Silicon detectors and
compatible with medium-fast scintillator like NaI, CsI, BGO
Software selectable coarse gain (x1, x2, x4, x8)
Jumper selectable x0.2 attenuation
10 μs AC coupling jumper selectable for TRP detectors
Digital oscilloscope for an easy acquistion parameters setup and
signal monitoring
Features Digital Pulse Processing for PHA, providing Energy and
Timestamp in list mode
Software adjustable digital shaping filter, baseline restoration and
fine gain
Online pile-up rejection and software dead time evaluation
Suited also for high count rate application
Configurable coincidence or anti-coincidence of the input analog
signals for background rejection and anti-compton shield
VME and Optical Link communication interface
Drivers, libraries and API for Windows and Linux 32/64 bit
MC2Analyzer and CoMPASS software to manage the acquisition and perform basic spectrum analysis
Overview
The V1782 is the new CAEN Octal 32k digital MCA. Designed for high energy resolution semiconductor detector is perfectly suited for those application in which the number of input channels start becoming relevant such as when segmented HPGe, Clovers and silicon detectors are involved.
The V1782 provides four steps of software selectable coarse gain and two possible jumper selectable dynamical ranges (0.2-0.4-0.8-1.6 Vpp and 1-2-4-8 Vpp).
It is also compatible with Transistor reset preamplifier thanks to the jumper selectable 10 us AC coupling.