Characteristics
- Measured entity
- water
- Application domain
- monitoring
- Measured value
- water hardness, water activity, purity, Alkalinity, frequency response
- Configuration
- compact, modular
- Mode of use
- automatic
- Technology
- colorimetric
- Protection level
- IP65
- Other characteristics
- continuous, digital, sampling, real-time, RS485, high-definition, dynamic, temperature control, ATEX
Description
Analysis of water hardness
(with good water/not good water signal)
Multifunctional microprocessor analyser for water hardness analysis with alarm set point for:
French degrees: 0,05 / 0,09 / 0,18 / 0,36 / 0,54 / 0,9 / 1,8 / 3,6 / 5,4 or 9,0 °f.
German degrees: 0,02 / 0,05 / 0,1 / 0,2 / 0,3 / 0,5 / 1,0 / 2,0 / 3,0 or 5,0 °dH
ppm CaCO3: 0,4 / 0,9 / 1,8 / 3,6 / 5,4 / 9,0 / 18,0 / 36,0 / 54,0 or 90,0 ppm CaCO3
mmol/l: 0,004 / 0,01 / 0,04 / 0,50 / 0,09 / 0,18 / 0,36 / 0,54 or 0,90 mmol/l
carbonate hardness with alarm set point for:
French degrees: 1,8 / 2,7 / 3,6 or 5,4 °f
German degrees: 1,0 / 1,5 / 2,0 or 3,0 °dH
The alarm set point is established by the indicator.
The DUROMAT® IQ system controls the softening system in a qualitative way, thus avoiding problems or damage caused by water hardness. This saves money: water, time and chemicals that would require the eventual early regeneration of the sweetening system.
You have a display with indication of hardness above or below the alarm set point (good water/not good water)
Display: fresh water, hard water, analysis in force, repetition analysis (due to hardness exceeding), analysis stopped, pre-alarm indicator level <10%, lack of indicator (100% exhausted), appliance malfunction.
SD card for saving the results.
Button INFO: last 100 results with date/time
At the output we have three relays for the signal of exceeding hardness and malfunction alarm.
An entry for STOP or START analysis.
Buttons for: start of analysis, measurement chamber cleaning, relay reset and manual dosage.
Programmable analysis interval.
You can repeat an analysis that is not good (just exceeded the maximum value, without giving an alarm signal).